69
E9000
HANDPIECES
B
E9414 Tubing Set with Recirculation only:
PRECAUTION: When using the
E9414 Tubing Set, the tubing line
MUST be fully primed (no air in line)
or the handpiece will overheat, which will
cause damage to the handpiece.
1. Prime the tubing by doing the following:
(a) Squeeze the priming chamber.
(b) While continuously squeezing the
priming chamber, operate the drill at
low speed until fluid completely fills
the tubing and is recirculated back into
the fluid bag. Ensure fluid remains in
the priming chamber during use with
no air in the tubing.
2. Snap the handpiece cord clips onto the
handpiece cable to secure the tubing.