100/120 VAC:
Use only a listed (UL, CSA) detachable power cord manufactured to the following specifications.
Plug End
NEMA 5-15P hospital grade, 15 amps, 125V
Receptacle End
IEC 320/CEE-22, 6 amps, 250V/15 amps, 125V
Cord
UL style SJT, 18 AWG, 3 conductor
220/240 VAC
Plug End
Molded straight PVC plug with double grounding system
DIN 49441, CEE 7/U11, 10/16A, 250V
CEBEC, DEMKO, KEMA, NEMKO, OVE, SEMKO, VDE, UTE, FEMKO
Receptacle End
Molded straight PVC plug
DIN 49457, CEE 22/V, 10A, 250V
VDE, D, N, S, SEV, OVE, KEMA
Cord
PVC, 7.2mm diameter
10A, 250V
Conductors: 3 x 1 mm
2
Conductor Colors - brown, blue, green/yellow stripe