126
MAINTENANCE
F
4.2 Troubleshooting
Symptom Possible Cause Corrective Action
Controller does not
power on when
standby power switch
is turned on.
Power cord unplugged or faulty.
Blown fuse(s).
Plug in power cord.
Replace power cord.
Replace fuses. Reference “4.4 Fuse
Replacement and Voltage Selection
Means” on page 130.
Handpiece does not
operate.
Handpiece cord not connected
securely, if applicable.
Handpiece cord and/or button, if
applicable, faulty.
Handpiece faulty.
If using a footswitch, footswitch
cord not connected securely.
Footswitch cord or footswitch
faulty.
Handpiece safety is in the safe,
or off position, if applicable.
Securely connect handpiece cord to
the handpiece, if applicable, and the
controller receptacle.
Replace handpiece cord, where
applicable. If handpiece cord is
permanently affixed, return
handpiece for service.
Return for service.
Securely connect footswitch cord to
controller receptacle.
Return for service.
Move safety, if applicable, to the
appropriate operating position.