106
MINI-DRIVER
HANDPIECE
E
3.10.4 Sagittal Saw Attachment (K220)
NOTES:
1. The K220 Sagittal Saw attachment only
accepts K140 and K150 series blades.
2. For more precise osteotomies, lower
cutting temperatures, and reduced
instrument wear, use a new blade for
each procedure.
3. Sagittal Saw blades are single
use only. Dispose of properly
after use.
1. To attach the Sagittal Saw attachment,
reference “3.10.2 Connecting/Removing
Attachments” on page 102.
2. The Sagittal Saw attachment may be placed
in any of four positions at 90 degree angles.
3. To attach a blade:
(a) Ensure the handpiece is in the
safe, or off position before
attaching or removing blades.
(b) Insert the blade into the blade holder in
the desired blade position. Blades may
be placed in any of five positions with a
180 degree radius.
(c) Tighten the blade by inserting the K201
Wrench and turning clockwise until
secure.
(d) Ensure the blade is secured by pulling
outward on the blade.