EasyManua.ls Logo

ConMed Advantage - High Speed Shaver (E9005) Operation

ConMed Advantage
166 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
62
E9000
HANDPIECES
B
4. The handpiece cord may be disconnected
from one handpiece and connected to
another handpiece, if applicable, without
turning off the controller.
5. Handpieces are inoperable if the
activation lever or trigger is depressed
while connecting the handpiece cord.
6. The MicroChoice Modular Handpiece
(5020-027), Mini-Driver Handpiece
(K500), or any of the Power
Pro
handpieces will not operate by the
footswitch.
Group 1 Handpieces
NOTES:
1. Group 1 handpieces consist of the High
Speed Shaver (E9005), CoolFlex Drill
(E9010), and Cranial Perforator Drive
(E9015).
2. Group 1 handpieces will operate on the
E9000 controller and only in Port 2 (Top
Portion) of the Advantage controller.
3.1 High Speed Shaver (E9005)
Operation
NOTE: If irrigation is being utilized, the
E9416 (4.2 mm) or E9417 (3.7 mm) Tubing
Set must be used.
1. To insert a shaver blade or bur:
(a) Open the collet by turning the locking
collet counterclockwise as viewed
from the collet end of the handpiece.
(b) Insert a shaver blade or bur until the
shaft is completely seated.
(c) If using a shaver blade, rotate the
shaver blade until the cutting window
is in the desired position.
(d) Secure the shaver blade or bur by
turning the locking collet clockwise.

Table of Contents