122
MAINTENANCE
F
4.1.4 Attachment Lubricating
Instructions
PRECAUTIONS:
1. Only the 20° (5020-063, 5020-064),
70° (5020-065, 5020-069), and 90°
(5020-066) angle attachments, the
following Low Speed Drill Attachments
(5020-034, 5020-035), and the
MicroChoice Wiredriver Attachment
(5020-028) are to be lubricated. Do not
lubricate any handpieces or other
attachments.
2. Attachments must be cleaned prior to
lubrication.
3. Use only Pana Spray Attachment
Lubricant (1375-037).
WARNING: The spray container is
pressurized and the contents are
flammable (contains LP gas). Do not
expose can to temperatures exceeding 40°C
(104° F). Do not puncture or burn can, even
when empty. Do not spray on or near open
flames or incandescent material.
1. Before using, shake the can well. Read all
the information and instructions on the
spray can label.
2. If the spray nozzle is not attached, securely
attach it to the spray can dispenser.
3. Completely insert the spray nozzle in the
bottom of the attachment.
4. With the can in an upright position, depress
the button and spray the lubricant for one to
two seconds or until lubricant flowing from
the attachment is clear and free of traces of
blood. Wipe any excess lubricant from the
attachment.
5. After lubrication is complete, connect the
attachment to the handpiece and operate for
approximately 15 seconds.