121
MAINTENANCE
F
4. With the nose of the handpiece facing
downward, thoroughly scrub the handpiece,
handpiece cord, attachments, and activation
lever, if applicable, with a clean, soft brush
dampened with a mild, pH-balanced
detergent. Remove all traces of blood,
debris and stains. Do Not immerse
equipment in soap solution or rinse water.
5. To clean the cannulated section of the
MicroChoice Modular, and Mini-Driver
handpieces:
(a) If using a wire guard, if applicable,
remove it from the handpiece.
(b) Feed the wire end of a cleaning brush
through the back of the handpiece.
(c) Pull the brush through the handpiece.
Repeat until all debris is removed.
6. Manipulate all moving parts of the
handpiece and attachments to ensure all
debris is removed. If not, clean again until
all debris is removed.
7. Keeping the nose of the handpiece pointed
downward, rinse under running water to
remove all traces of soap. Rinse all
attachments likewise.
8. Flush all surfaces free of tap water with
distilled water to prevent metal
discoloration.
9. Gently shake the equipment free of water
and wipe the surfaces with a clean, lint-free
towel.