MAINTENANCE
F
4. With the nose of the handpiece facing
downward, thoroughly scrub the handpiece,
handpiece cord, and guards with a clean,
soft brush dampened with a mild, pH-
balanced detergent. Remove all traces of
blood, debris and stains. Do Not immerse
equipment in soap solution or rinse water.
(a) On the Cranial Perforator Drive (E9015)
thoroughly clean the chuck by pulling
the chuck sleeve back and scrubbing the
area with a soft bristled brush.
(b) On the CoolFlex High Speed Drill,
thoroughly clean the collet area with a
soft bristled brush.
(c) Using a cleaning brush, thoroughly
clean the inside of the bur guard.
Repeat until all debris is removed.
5. Manipulate all moving parts of the
handpiece to ensure all debris is removed. If
not, clean again until all debris is removed.
6. Keeping the nose of the handpiece pointed
downward, rinse under running water to
remove all traces of soap.
7. Flush all surfaces free of tap water with
distilled water to prevent metal
discoloration.
8. Shake the equipment free of water and wipe
the surfaces with a clean, lint-free towel.
4.1.3.3 MicroChoice Drills, Saws,
Modular Handpiece, and Mini-
Driver Handpiece Cleaning
Instructions
1. Remove all attachments (i.e., chucks, saws)
and accessories (i.e., saw blades, burs) from
the handpiece prior to cleaning. For levered
handpieces, remove the Activation Lever
(Reference “3.5.4 Leverless Handpiece
Operation” on page 87 for removal
information).
2. Disconnect the handpiece cord from the
controller prior to cleaning.
3. Ensure the handpiece cord is securely
fastened to the handpiece during cleaning.