119
MAINTENANCE
F
4.1.3.1 MicroChoice, Advantage, and
APEX Shaver Handpiece
Cleaning Instructions
NOTE: The following cleaning instructions
apply to the following MicroChoice, APEX,
and Advantage Shaver Handpieces: MC9820/
C9820/D9820/D9920 Basic Handpieces,
C9824/D9824/D9924 2-Button Handpieces,
MC9828/C9828 Full-Function Handpieces,
and MC9840/C9840 Small Shaver
Handpieces.
1. Disconnect the handpiece cord from the
controller prior to cleaning.
2. Remove all accessories (i.e., burs, shaver
blades) from the handpiece prior to
cleaning.
3. Ensure the suction control valve is in the
“ON”, or forward position prior to cleaning.
4. This step applies ONLY to the MC9820
Basic, MC9828 Full-Function, and
MC9840 Small Shaver handpieces (older
version with removable handpiece cords):
(a) Remove the handpiece cord from the
handpiece prior to cleaning and/or
immersing in water for cleaning.
5. Thoroughly scrub the handpiece and
handpiece cord with a clean, soft brush
dampened with a mild, pH-balanced
detergent. Remove all traces of blood,
debris and stains. Do Not immerse
equipment in soap solution or rinse water,
other than the handpieces indicated
previously in step 5.
6. Clean the suction tube thoroughly with a
cleaning brush until all debris is removed.
(a) Feed the wire end of the cleaning brush
through the back of the handpiece.
Repeat until all debris is removed.
7. Manipulate all moving parts of the
handpiece to ensure all debris is removed. If
not, clean again until all debris is removed.
8. Rinse under running water, or immerse the
handpieces that can be immersed as
indicated previously in step 4, to remove all
traces of soap.
9. Flush all surfaces free of tap water with
distilled water to prevent metal
discoloration.
10. Gently shake the equipment free of water
and wipe the surfaces with a clean, lint-free
towel.
4.1.3.2 E9000 Handpiece Cleaning
Instructions
NOTE: The following cleaning instructions
apply to the following handpieces: E9005,
CoolFlex, and E9015.
1. Remove all attachments (i.e., bur guards)
and accessories (i.e., bits, burs) from the
handpiece prior to cleaning.
2. Disconnect the handpiece cord from the
controller prior to cleaning.
3. Ensure the handpiece cord is securely
fastened to the handpiece during cleaning.