105
MINI-DRIVER
HANDPIECE
E
(d) Loosen the adjustment sleeve one half
turn. The pin or wire should slide freely
within the Pin/Wire Driver until the
grasping lever is depressed.
(e) For pin/wire sizes that fall between the
pin/wire size graduation, loosen the
adjustment sleeve one full turn. The pin
or wire should slide freely within the
Pin/Wire Driver until the grasping lever
is depressed.
2. To operate the handpiece:
(a) Release the safety and place the
handpiece in the forward position.
(b) To grip and drive the wire or pin,
depress the grasping lever until flush
with the handpiece and depress the
trigger.
3. To reposition the handpiece on the wire or
pin:
(a) Release the trigger and grasping lever.
(b) Slide the handpiece along the wire or
pin.
(c) Follow step 2 to further drive the wire or
pin.
4. To remove threaded wires from the patient:
(a) Insert the wire into the front of the
attachment (see step 1(a) through 1(e)).
(b) Place the handpiece in the reverse
position.
(c) Simultaneously squeeze the grasping
lever and depress the trigger.