127
MAINTENANCE
F
The CoolFlex High
Speed Drill overheats.
Tubing set is not completely
primed.
Prime tubing set by following the
priming procedure (See “ E9414
Tubing Set with Recirculation
only:” on page 69.
A MicroChoice, Group
2 handpiece overheats.
Worn bearings in the bur guard.
Moisture in the handpiece.
Reference “4.5 Bur Guard Testing
Procedure” on page 131. If
resistance is noticed, return bur guard
for service. If no resistance is
noticed, return handpiece for service.
Resterilize the handpiece according
to sterilization parameters and drying
times, reference “4.1.5 Sterilization
Information” on page 123.
Leakage occurring at
the tubing handpiece
cassette and handpiece
interface.
Handpiece cassette not
completely or incorrectly
attached to the handpiece.
Remove the tubing handpiece
cassette from the handpiece and
reinstall. If leakage still occurs,
replace with a new tubing set.
Symptom Possible Cause Corrective Action