129
MAINTENANCE
F
Error Messages (Continued)
“STALL CHECK
BUR LOCK
message displays upon
initial activation of the
CoolFlex High Speed
Drill while a bur is
locked in place.
The drill was not cleaned
correctly and the bearings are
locked due to debris build-up.
Disconnect the handpiece from the
controller. Rotate the bur in the
handpiece with your fingers until the
bur spins freely. If the message still
displays when the handpiece is
activated with the footswitch, return
the handpiece and bur guard to
ConMed Linvatec for service.
Magnetic Field
is displayed on the
controller.
The handpiece is in close
proximity to a magnetic drape.
Removable activation lever
partially connected.
Faulty handpiece.
Remove the handpiece from the
magnetic drape area.
Completely connect the activation
lever (See “3.5.4 Leverless
Handpiece Operation” on page 87).
Return handpiece to ConMed
Linvatec for service.
CONNECT HAND-
PIECE
message displays
when the handpiece is
already plugged into
the controller.
Handpiece or handpiece cord not
connected securely.
Handpiece, handpiece cord and/
or controller faulty.
Securely connect handpiece or
handpiece cord.
Replace handpiece. If symptom is
corrected, return handpiece for
service.
Else, replace handpiece cord. If
symptom is corrected, return faulty
handpiece cord.
Else, return controller for service.
Symptom Possible Cause Corrective Action
N
S
PowerPro