148
TECHNICAL
SPECIFICATIONS
G
Description Duty Cycle at 25°C Ambient
Controller (D3000, D3000I, E9000, E9000G) Designed to operate at its rated maximum
load of 325 VA output for a time period of
2 minutes ON, 6 minutes OFF.
Controller (D3000J, E9000J) Designed to operate at its rated maximum
load of 325 VA output for a time period of
2 minutes ON, 8 minutes OFF.
High Speed Shaver Handpiece (E9005) 30 seconds ON, 5 minutes OFF.
CoolFlex High Speed Drill Handpiece (E9010) 30 seconds ON, 6 minutes OFF.
Cranial Perforator Drive (E9015) 1 minute ON, 3 minutes OFF.
MicroChoice Medium Speed Drill (5020-021) 2 minutes ON, 3 minutes OFF.
MicroChoice Sagittal Saw (5020-022) 30 seconds ON/OFF for 6 cycles, then wait
30 minutes.
MicroChoice Reciprocating Saw (5020-023) 30 seconds ON/OFF for 4 cycles, then wait
30 minutes.
MicroChoice Oscillating Saw (5020-024) 30 seconds ON/OFF for 4 cycles, then wait
30 minutes.
MicroChoice Elite High Speed Drill (5020-025) 2 minutes ON, 8 minutes OFF.
MicroChoice Low Speed Drill (5020-026) with Drill
Head
1 minute ON, 3.5 minutes OFF.
MicroChoice Low Speed Drill (5020-026) with Screw/
Tap Head
100% duty cycle.
MicroChoice Modular Handpiece (5020-027) 100% duty cycle.
MicroChoice Small Shaver Handpiece (MC9840) 1 minute ON, 3 minutes OFF (no irrigation).
100% duty cycle (with irrigation).
Mini-Driver Handpiece (K500) 1 minute ON, 3 minutes OFF
Advantage Basic Shaver Handpiece (D9820) 100% duty cycle (with irrigation).
Advantage 2-Button Shaver Handpiece (D9824) 100% duty cycle (with irrigation).
Advantage Turbo Basic Shaver Handpiece (D9920) 100% duty cycle (with irrigation).
Advantage Turbo 2-Button Shaver Handpiece (D9924) 100% duty cycle (with irrigation).