Continuous tone – volume must remain constant for
30 seconds.
If not, reconnect to mains and leave machine on for
24 hours so that battery can recharge. Repeat test!
Switch off machine. Press power switch again and
turn to O. Alarm goes out.
Reconnect to power supply.
Self-test
The self-test must be completed successfully in order to
establish readiness for operation!
Supplementary equipment must not be switched on until
after the self-test.
Start self-test:
Switch on machine: press power switch.
Test proceeds as described from page 31 onwards.
Calibrate flow sensor, see page 30
Calibrate inspiratory O2 sensor with
100% O
2
This is only necessary if sidestream O2 measurement is
not used.
Remove O2 sensor and place test adapter 68 01 349
on sensor.
Allow an O2 flow of approx. 1 L/min to flow over the
O2 sensor for approx. 2 minutes.
Select »more« with the rotary control.
Select »O2 sensor 100 Vol.%« with the rotary control.
Display:
When the O2 sensor has been flushed with O2 for
approx. 2 minutes:
Press rotary control to confirm; calibration is then
started and continues automatically.
The clock icon is replaced by a tick () when calibration
is complete.
Replace the O2 sensor in its mount.
103
Check operational readiness
Self-test
Calibrate flow sensor
Calibrate inspiratory O
2
sensor
Standby / Configuration Alarms inactive!
O
2
sensor
100 vol. %
linearity
test O
2
CO
2
sensor
Detach O
2 sensor and
flush with 100 vol. % O
2.
calibrating
O
2
sensor
21 vol.%
flow sensor
Ventilator-
start up test
more
Calibrate O
2 sensor after
replacing sensor and after 24 hrs.
O2
ca. 1 L/min
07228971 19428970