87
Care
Care methods
Tabular summary
Tabular summary
Method
Disinfection by Cleaning with Formaldehyde Steam sterilization
Machine components wiping or immersion automatic cleaners vapour at 121 °C at 134 °C
Stripped down basic machine Wipe ●
Parameter box (optional) Wipe
Rubber parts:
Tube / mask ●● ●●
Y-piece ●● ●●
Breathing hoses ●● ●●
Breathing bag ●● ●●
Breathing bag hose ●● ●●
Roller diaphragm of piston pump ●● ●●
Anaesthetic gas scavenging system ●● ●●
Pressure measuring hoses ●● ●●
Connecting hoses for measured
gas return system ●● ●●
Metal parts:
Cylinder head of piston pump ●●●
Piston of piston pump ●●●
Parts of breathing system ●●●
Valves of breathing system ●●●
Plastic parts:
Housing of piston pump ●● ●●
Soda line container with screen ●● ●●
T-piece for sidestream gas
measurement ●● ●●
Water trap for measured gas ●● ●●
Water traps for airway ●● ●●
Filters:
Microbial filter 24 times
Sensors:
Temperature sensor ●
(airway temperature)
Temperature sensor ●
(skin/rectal/oesophagus)
O2 sensor (housing) ●
Flow sensor externally
The following parts cannot be processed and prepared:
(must be replaced if damaged or possibly contaminated)
Filter of pressure measuring hose: Change every 6 months or when resistance increases (e.g. water absorption).
The warning message P-Sensor fault resp. VENT ERR !!! is output.
Filter of measured gas return hose: Replace every 6 months.
for sidestream gas measurement:
Water separator: Replace, depending on degree of contamination
Connecting hoses: Replace after each patient
Filter in T-piece: Replace after each patient