87
Care
Care methods
Tabular summary
Tabular summary
Method
Disinfection by Cleaning with Formaldehyde Steam sterilization
Machine components wiping or immersion automatic cleaners vapour at 121 °C at 134 °C
Stripped down basic machine Wipe
Parameter box (optional) Wipe
Rubber parts:
Tube / mask ●●
Y-piece ●●
Breathing hoses ●●
Breathing bag ●●
Breathing bag hose ●●
Roller diaphragm of piston pump ●●
Anaesthetic gas scavenging system ●●
Pressure measuring hoses ●●
Connecting hoses for measured
gas return system ●●
Metal parts:
Cylinder head of piston pump ●●
Piston of piston pump ●●
Parts of breathing system ●●
Valves of breathing system ●●
Plastic parts:
Housing of piston pump ●●
Soda line container with screen ●●
T-piece for sidestream gas
measurement ●●
Water trap for measured gas ●●
Water traps for airway ●●
Filters:
Microbial filter 24 times
Sensors:
Temperature sensor
(airway temperature)
Temperature sensor
(skin/rectal/oesophagus)
O2 sensor (housing)
Flow sensor externally
The following parts cannot be processed and prepared:
(must be replaced if damaged or possibly contaminated)
Filter of pressure measuring hose: Change every 6 months or when resistance increases (e.g. water absorption).
The warning message P-Sensor fault resp. VENT ERR !!! is output.
Filter of measured gas return hose: Replace every 6 months.
for sidestream gas measurement:
Water separator: Replace, depending on degree of contamination
Connecting hoses: Replace after each patient
Filter in T-piece: Replace after each patient