Monitoring with reduced alarm limits
The »IPPV alarm limits« are automatically activated on
the monitor when a ventilation frequency equal to or
greater than 6 breaths per minute is set in »IPPV« and
»SIMV« modes.
The »SIMV alarm limits« are automatically activated
on the monitor if a ventilation frequency of less than
6 breaths per minute is set in »SIMV« mode.
The »MAN/SPONT alarm limits« are automatically
activated on the monitor in »MAN/SPONT« mode. This
prevents false alarms (see »Alarm concept«, page 65
onwards).
Checking water traps
WaterLock
(refer also to the separate Instructions for Use)
When fouled
or
if CO2 LINE ? ! is output although the sample line is clear.
● Replace water trap:
– when severely fouled,
– if the error message persists even after draining the
water trap,
– when the maximum service period of four weeks is
exceeded.
● Grasp the ribbed sides of the water trap and draw it
out.
● Remove a new water trap from the package.
● Enter the date in the field provided on the water trap.
● Grasp the ribbed sides of the water trap and slide it
into the holder – until it tangibly engages.
● Empty water trap:
● Insert an empty syringe without cannula, at least
20 mL, into the connector.
● Extract the water, remove syringe and dispose of the
full syringe as ordinary domestic waste.
● Push water trap back into holder – until it tangibly
engages.
Disposal:
● Water traps which have been dismantled must be
disposed of as ordinary commercial waste.
Waste code number 91101.