EasyManua.ls Logo

Dräger Cato - C-Lock ECG Synchronization; Operating Parameters

Dräger Cato
148 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
121
Technical data
Monitor
Operating parameters
C-Lock ECG synchronization (optional)
Prerequisite for the
ECG synchronization signal ......: Pos. pulse with > 4.5 V and > 10 ms duration to drive 2 mA.
Max. permitted signal delay
in relation to the current
QRS complex ...........................: 40 ms
Signal isolation from
remaining electronics ................: Dielectric strength > 4 kV
Socket for 2-pin pawl-type
connector, dia. 3.5....................: Pin assignment:
Operating parameters
Noise emission ..................................... : Max. 56 dB(A) (as determined in a free field over a reflecting surface)
Protection classes:
SpO2 sensor ....................................... : Type BF electrically isolated from protective earth conductor
Electromagnetic compatibility ............... : Tested to EN 60601-1-2
Classification ...................................... : Class II b
as per Directive 93/42/EEC
Annex IX
UMDNS-Code ..................................... : 10-134
Universal Medical Device
Nomenclature System
Technical documentation available on request
m
signal
ground
07728971

Table of Contents