Power supply
Equipotential bonding conductor connected?
Mains plug connected?
1 Switch on equipment –
Press main switch to I.
Self-test
Check the signals and tick off the following on the
checklist:
Lamp and alarm test, in succession:
Ventilator:
Indication of the software version, all indicators light
up for approx. 2 seconds while a single tone sounds.
Monitor:
The machine runs through the self-test:
All LEDs and display elements light up for
approx. 2 seconds. The LED in the Standby key E
continues to light up.
Two alarm tones sound.
The internal program memories are tested.
The self-test is completed after approx. 1 minute.
The following display appears on the monitor:
»Self-test«
is displayed on the ventilator.
If the self-test reveals a fault of no relevance to the
safety of the equipment and not affecting any of the
measuring functions, the following message appears on
the monitor:
»Ready within limits«
together with a specific error message (refer to
pages 74 onwards). However, the equipment can be
operated nevertheless.
Press the rotary control.
Call DrägerService.
26
Preparing for use
Checking readiness for operation with checklist
Power supply
Self-test
1
Technology for life
Cato
D
SW-Version 2.05
02-05-00
self-test
00137225 0113722404428970