Dismantling parts
T-piece of measured gas hose (if used)
1 Remove filter.
Inspiratory O2 sensor (if used)
Must not be disinfected in liquid or autoclaved!
Wipe any dirt off housing or lead with damp
disposable cloth.
Unscrew O2 sensor and remove O2 cell. Do not
touch wire mesh of O2 cell, otherwise it loses its
water-repellent property and measurement becomes
inaccurate.
Dirt on wire mesh of sensor cell is simply wiped off
with a disposable cloth moistened with distilled water.
Temperature sensor (breathing gas) (optional)
Draw temperature sensor out of Y-piece, remove
Y-piece from ventilation hoses.
Prise lead out of hose clips.
Pull connector out of socket.
Wipe dirt off with damp disposable cloth.
Flow sensor
The flow sensor cannot be disinfected or sterilized.
Contamination by infectious patients: dispose of as
directed!
If contamination is unlikely, the flow sensor can be
reused as long as it can be calibrated correctly.
81
Care
Dismantling parts
T-piece of measured gas hose
Inspiratory O
2
sensor
Temperature sensor, Flow sensor