Dismantling parts
T-piece of measured gas hose (if used)
1 Remove filter.
Inspiratory O2 sensor (if used)
● Must not be disinfected in liquid or autoclaved!
● Wipe any dirt off housing or lead with damp
disposable cloth.
● Unscrew O2 sensor and remove O2 cell. Do not
touch wire mesh of O2 cell, otherwise it loses its
water-repellent property and measurement becomes
inaccurate.
Dirt on wire mesh of sensor cell is simply wiped off
with a disposable cloth moistened with distilled water.
Temperature sensor (breathing gas) (optional)
● Draw temperature sensor out of Y-piece, remove
Y-piece from ventilation hoses.
Prise lead out of hose clips.
● Pull connector out of socket.
● Wipe dirt off with damp disposable cloth.
Flow sensor
The flow sensor cannot be disinfected or sterilized.
Contamination by infectious patients: dispose of as
directed!
If contamination is unlikely, the flow sensor can be
reused as long as it can be calibrated correctly.
81
Care
Dismantling parts
T-piece of measured gas hose
Inspiratory O
2
sensor
Temperature sensor, Flow sensor