Checking readiness for operation
with checklist
It is assumed that the following conditions have
been met:
The equipment and its accessory parts have been
cleaned and disinfected –
refer to page.......................................... 83 onwards
The unit is fully equipped for the application in
question –
refer to page.......................................... 89 onwards
The user has a good knowledge of the Instructions for
Use and has been trained to use the equipment.
An emergency ventilation bag with appropriate mask is
available on the equipment.
A checklist is affixed to the equipment.
The equipment must always be checked against the
checklist before it is used!
This check is mandatory as specified by EN 740!
Duration: approx. 5 minutes (depending on the scope
of calibration).
The equipment is subsequently in »standby« mode
when ready for use. Anaesthesia ventilation can
immediately be started at any time in this mode –
refer to page.......................................... 36 onwards
Update the checklist by adding or deleting points
in accordance with the Instructions for Use, the
equipment type and configuration concerned and the
various supplementary units connected. It will then
contain all the requisite checks.
Enter the model designation and serial number of the
equipment.
Tick off the results of the checks in the ACTUAL
column.
Remember to sign and date the checklist!
Test sequence
Note any alterations and additions!
Note the Instructions for Use of the individual units!
If the checks do not proceed as planned, restore the
required status.
21
Preparing for use
Checking readiness for operation with checklist