Breathing system
Lift off table top:
1 Lever set to the position shown in black.
The following are complete:
2 Hose with manual ventilation bag Symbol:
(Connected from below – not shown)
Correct breathing hoses installed.
(Adult or infant hoses)
3 Pressure measuring hose with filter connected.
4 Measured gas return hose Symbol:
connected.
5 Fresh gas hose plugged in.
(Connected from below – not shown)
6 Valve discs inserted.
Pressure limiting valve (APL) present.
Inspiratory microbial filter. Symbol:
Expiratory microbial filter. Symbol:
Soda lime
7 Lime has not noticeably changed colour (purple).
Filling level adequate (up to the mark).
Soda lime container is securely tightened – up to the
end stop (clockwise).
Emergency ventilation bag
(not shown)
Bag is complete with mask and hung from the side
of the Cato.
Bag functions correctly.
Check and tick off.
23
Preparing for use
Checking readiness for operation with checklist
Breathing system, Soda lime
Emergency ventilation bag
1
275
APL
3
4
6
035289700362897003728970