Technical data
Gas delivery
Monitor
118
Gas delivery
Low-flow measuring tubes ........: (Calibrated for 20 °C, 1013 mbar)
O2: 0.02 to 0.5 L/min ± 10 %
At minimum scale: + 20 %, – 10 %
At minimum scale: ± 5 %
O2: 0.55 to 10.0 L/min ± 10 %
At minimum scale: + 20 %, – 5 %
N2O: 0.02 to 0.5 L/min ± 10 %
At minimum scale: 20 %, + 10 %
At minimum scale: ± 5 %
N2O: 0.55 to 10.0 L/min ± 10 %
At minimum scale: + 20 %, – 5 %
AIR: 0.2 to 14.0 L/min ± 10 %
At minimum scale: + 15 %, – 5 %
Delivery output..........................: O2, N2O, AIR: at least 9 L/min each
Monitor
Pressure measurement
Airway pressure........................: 10 to 80 mbar
Resolution ................................: 1 mbar
Accuracy ..................................: at least 1 mbar or better than ± 4% of the measured value
Flow-Messung
Tidal volume VT.........................: Range: 0.01 to 9.99 L
Resolution: 0.01 L
Accuracy: better than 0.01 L or ±8%
1)
of the measured value
(under calibration conditions and 1013 hPa)
Minute ventilation AMV .............: Range: 0 to 99.9 L/min
Resolution: 0.1 L/min
Accuracy: better than 0.2 L/min or ±8%
1)
of the measured value under
calibration conditions and BTPS (BTPS = Body Temperature,
present atmospheric Pressure, 100% humidity Saturated)
Respiration rate ........................: Range: 0 to 80 per minute
Resolution: ± 1 per minute
Accuracy: ± 1 per minute
1) When using an microbial filter on the expiratory side, the tolerance
increases to ±15%.
O2 measurement in main stream (inspiration line)
Range.......................................: 5% to 100% by volume
Resolution ................................: 1% by volume
Accuracy ..................................: calibrated with air: ±3 % by vol. in 5 to 50% by vol. measuring range
±5 % by vol. in 50 to 100% by vol. measuring range
calibrated with 100 % O2: ±3% in the 5 to 100% by vol. measuring range
Response time t
10...90
...............: better than 15 seconds