119
Technical data
Monitor
O2, CO2 and anaesthetic agent measurement in sidestream (sampling)
Sampling rate (selectable).........: 200 mL/min
60 mL/min
O2 measurement
Range.......................................: 5% to 100% by volume
Resolution for FiO2 and FeO2...: 1% by volume
Resolution for ∆O2....................: 0.1% by volume
Accuracy ..................................: calibrated with air: ±3 % by vol. in 5 to 50% by vol. measuring range
±5 % by vol. in 50 to 100% by vol. measuring range
calibrated with 100 % O2: ±3% in the 5 to 100% by vol. measuring range
Response time t
10...90
...............: At 200 mL/min: better than 500 ms
At 60 mL/min: better than 1 s
CO2 measurement
Special feature .........................: When the CO2 alarm is activated, the inspiratory CO2 concentration in the
breathing air is measured with a fixed upper limit of 5 mmHg)
Measuring range.......................: 0 to 9.9% by volume (corresponding to 0 to 9.9 kPa or 0 to 80 mmHg)
Accuracy ..................................: ± 0.2 % by volume or 5% of the measured value
Resolution ...............................: 0.1% by volume (0.14 Pa or 1 mmHg)
Response time t
10 ...90
..............: at 200 mL/min: 300 ms
at 60 mL/min: 1 s
Warm-up phase........................: 4 minutes (to achieve ISO/DIS 11196 accuracy level)
2 minutes (for first measured values)
Zero drift ..................................: Within the specified accuracy, without limitation in time
Anaesthetic agent measurement
Display range for N2O ..............: 0% to 100% by volume
Accuracy: 0 to 40% better than ± 2.5% by volume absolute
41 to 100% better than ± 6.0% by volume absolute
Resolution: 1 % by volume absolute
Display range for halothane.......: 0% to 7.5% by volume
Accuracy: 0 to 5% better than ± 0.2% by volume absolute
5 to 7.5% better than ± 0.3% by volume absolute
Resolution: 0.1 % by volume absolute
Display range for
enflurane and isoflurane ............: 0% to 7.5% by volume
Accuracy: 0 to 5% better than ± 0.2% by volume absolute
5 to 7.5% better than ± 0.3% by volume absolute
Resolution: 0.1 % by volume absolute
Display range for desflurane......: 0% to 20% by volume
Accuracy: 0 to 10% better than ± 0.4% by volume absolute
10 to 20% better than ± 0.8% by volume absolute
Resolution: 0.1 % by volume absolute