Calibrate O2 sensor for sidestream
measurement with 100% O
2
This is necessary for the monthly linearity check.
Prepare a substitute sampling line:
Cut the sampling hose through the middle.
Unscrew the original sampling hose from the water
trap and screw on the slit sampling hose.
On Cato:
Disconnect fresh gas hose from breathing system.
Set an O2 flow of 1 L/min O2 at the O2 flow tube and
slide the sampling hose into the fresh gas hose.
Use the rotary control to select »calibrating« and then
»more«.
Select »O2 sensor 100 Vol.%« with the rotary control.
Display:
Allow O2 to flow for approx. 2 minutes.
Press rotary control to confirm; calibration is then
started and continues automatically. The clock icon
() is then replaced by a tick (
u
).
Screw the original sampling hose back onto the
water trap.
Plug the fresh gas hose back into the breathing
system.
Check linearity
Must be checked every month.
First calibrate the O2 measurement used (inspiratory or
sidestream) with 100% O2 (see above).
Then:
Expose the inspiratory O2 sensor to ambient air for
approx. 2 minutes or,
in the case of the O2 sensor for sidestream measurement:
unscrew the sampling hose from the water trap and
allow it to take in air for approx. 2 minutes.
Display on screen should show between
18 and 24 vol.% O2.
If value displayed is outside
18 to 24 vol.% O2
sensor capsule is faulty.
Check operational readiness
Calibrate O
2
sensor for sidestream measurement
Check linearity
104
Standby / Configuration Alarms inactive!
O
2
sensor
100 vol. %
linearity
test O
2
CO
2
sensor
Expose end of sample line
to pure O
2.
confirm !
calibrating
O
2
sensor
21 vol.%
flow sensor
Ventilator-
start up test
more
Standby / Configuration Alarms inactive!
First calibrate with
100 vol.% O
2,
then expose sensor
to ambient air for about
2 minutes
O
2
21 vol. %
If the displayed value is outside
the range 18 to 24 vol.% O
2
,
the sensor is defective.
calibrating
O
2
sensor
21 vol.%
flow- sensor
Ventilator-
start up test
more
After fast, confirm
O
2
sensor
100 vol. %
linearity
test O
2
CO
2
sensor
0732897107428971