EasyManua.ls Logo

Dräger Cato - Manual Calibration of the Flow Sensor

Dräger Cato
148 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Manual calibration of the flow sensor
The flow sensor can be calibrated by hand while O2
calibration is still in progress, either in the machine or
after being removed from the machine.
Without removal:
On the anaesthetic unit:
1 Set AIR/N2O selector to »AIR«.
2 Close the delivery valves for O2 and N2O, open the
delivery valve for air and thoroughly flush the
breathing system with air.
3 Close the delivery valve for air.
Press the rotary control on the monitor to start
calibration.
The question mark (?) disappears and is replaced by
a timer icon (
u
) on the screen.
A tick ( ) appears on the screen when calibration
has been completed correctly.
Tick off in checklist.
The cursor frame has automatically jumped to
»Ventilator start-up test«.
With removal:
4 Remove flow sensor:
Unscrew expiration nozzle
Remove flow sensor
Briefly swing it to flush with ambient air
Hold it horizontally with the cable connection pointing
downwards (calibration in installation position
improves the measuring accuracy)
Seal off one or both sides as shown on the right,
preferably with your thumb or palm.
Press the rotary control on the monitor to start
calibration; continue as above.
4 Replace flow sensor!
30
Preparing for use
Checking readiness for operation with checklist
Manual calibration of the flow sensor
4
2
1
3
2
013372240512897005228970

Table of Contents

Related product manuals