EasyManua.ls Logo

Dräger Cato - Calibration

Dräger Cato
148 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Calibration
This menu is used for the following tests and calibrations:
Calibrate O2-sensor with 21 Vol.% O2
Calibrate flow sensor
Perform Ventilator start up test
more
The following can be realized under »more«:
–O2 sensor calibration with 100% O2 by volume,
linearity check of the O2 sensor and
calibration of the CO2 sensor.
The significance of the symbols is as before:
? = Enquires whether an action has been performed
or a setting made.
u
= Waiting period. The selected test step is being
carried out by the system.
= The action has been completed successfully or
is not required.
55
Monitor functions
Configuring monitor functions in »standby«
Calibration
Standby / Configuration
Alarms inactive!
calibrating
O
2
-sensor
21 Vol.%
flow sensor
Ventilator-
start up test
more
Calibrate O
2
sensor after
replacing sensor and after 24 hrs.
Detach O
2
sensor and expose
for 2 min. to ambient air.
confirm !
Standby / Configuration
Alarms inactive!
calibrating
O
2
sensor
21 Vol.%
flow sensor
Ventilator-
start up test
more
Only necessary if the automatic
calibration fails (INOP instead
of values).
With the flow sensor in place,
flush with air. Close flowtubes or
remove flowsensor and hold hori-
zontally with both ends closed.
confirm !
Standby / Configuration
Alarms inactive!
calibrating
O
2
sensor
21 Vol.%
flow sensor
Ventilator-
start up test
more
Test ventilator after system
start up.
Not possible while operational
?
038289710392897104028971

Table of Contents

Related product manuals