Changing patients
Changing parts
Refer to page 83 onwards for a schematic overview and
description of methods for cleaning and disinfection.
Maintenance intervals: page 106.
Machine in »standby« mode –
Vapor: control dial set to »0« = Off.
The T-piece and filter are not required when using a
Y-piece with Luer lock.
After treating an infectious patient
the entire machine must be cleaned, disinfected and
sterilized.
Note hygiene regulations –
risk of infection!
After treating a non-infectious patient
the following parts must be replaced before continuing
with the next patient:
Tube or mask
Y-piece
Both breathing hoses
Temperature sensor and cable if applicable
1 T-piece of the measured gas hose and filter, if used
(disposable articles, household refuse)
2 Drain:
Insert an empty syringe without cannula, at least
20 mL, into the connector.
Extract the water, remove syringe and dispose of the
full syringe as ordinary domestic waste.
Push water trap back into holder – until it tangibly
engages.
An overfull container or a faulty water trap could
lead to failure in sidestream measurement.
The following must also be changed when working
without microbial filter:
3 O2 sensor with connecting lead, if used.
4 Flow sensor.
Disconnect lead from flow sensor, it remains on the
machine.
Breathing bag with hose
Breathing system:
Disconnect fresh gas hose, it remains on the machine.
Pressure measuring hose and filter: must be replaced.
Hose connector remains on the machine.
Soda lime container – turn anticlockwise and pull
down – must be replaced.
Anaesthetic gas scavenging hose – release (latch in
front of connector) and pull off, also at the central
supply point.
47
Operation
Changing patients
1
2
4
G
A
S
3
[
08728970021372240892897009028970