TÜV Rheinland Energie und Umwelt GmbH
Luftreinhaltung
Report on the performance testing of the Serinus 30 ambient air quality moni-
toring system manufactured by Ecotech Pty Ltd measuring CO, Report no.:
936/21221977/D_EN
EP_Ecotech_Serinus30_en.docx
7.6 Detailed presentation of test results
Individual results are provided in Table 26.
Table 26: Individual results for the difference between sample and calibration port
Time Device 1 [µmol/mol] Device 2 [µmol/mol]
17:50 64.8 64.3
17:51 64.8 64.3
17:52 64.8 64.3
Average 64.8 64.3
17:59 64.9 64.4
18:00 64.9 64.4
18:01 64.9 64.4
Average 64.9 64.4
Deviation [%] -0.09 -0.17
Test gas to the sample port
Test gas to the calibration port