EasyManua.ls Logo

Ecotech Serinus 30 - Sensitivity Coefficient of Surrounding Temperature

Ecotech Serinus 30
254 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
TÜV Rheinland Energie und Umwelt GmbH
Luftreinhaltung
Report on the performance testing of the Serinus 30 ambient air quality moni-
toring system manufactured by Ecotech Pty Ltd measuring CO, Report no.:
936/21221977/D_EN
Page 59 of 253
EP_Ecotech_Serinus30_en.docx
6.1 5.3.7 Sensitivity coefficient of surrounding temperature
The sensitivity coefficient of surrounding temperature at zero and reference point
shall not exceed the requirements of Table 2 of VDI 4202, Sheet 1 (September
2010). A value c
t
at 70 % to 80 % of the upper limit of the certification range shall be
used as reference point.
The sensitivity coefficient of surrounding temperature shall not exceed
0.3 (µmol/mol)/K (i.e. (0.35 mg/)/K).
6.2 Equipment
Not applicable here.
6.3 Testing
Performance and evaluation of the steps taken to determine the sensitivity coefficient of sur-
rounding temperature are in line with the requirements stipulated in DIN EN 14626 (2012).
The reader is therefore referred to section 7.1 8.4.9 Sensitivity coefficient to the surround-
ing temperature.
6.4 Evaluation
Please refer to section 7.1 8.4.9 Sensitivity coefficient to the surrounding temperature.
6.5 Assessment
Please refer to section 7.1 8.4.9 Sensitivity coefficient to the surrounding temperature.
Does this comply with the performance criterion? yes
6.6 Detailed presentation of test results
Not applicable in this instance.

Table of Contents