TÜV Rheinland Energie und Umwelt GmbH
Luftreinhaltung
Seite 74 von 253
Report on the performance testing of the Serinus 30 ambient air quality
monitoring system manufactured by Ecotech Pty Ltd measuring CO, Report
no.: 936/21221977/D_EN
EP_Ecotech_Serinus30_en.docx
7.6 Detailed presentation of test results
Table 6: Individual readings for the response times for the component carbon monoxide
80%
0.0 0.9 1.0 1.0 0.1 0.0
0.00 62.
07 68.97 68.97 6.90 0.00
Cycle 1 t = 0 09:41:00 09:41:40 09:42:30 09:48:00 09:48:52 09:48:25
delta t 00:00:40 00:00:52
delta t [s] 40 52
Cycle 2 t = 0 09:54:00 09:54:56 09:55:10 10:00:00 10:00:52 10:01:29
delta t 00:00:56 00:00:52
delta t [s] 56 52
Cycle 3 t = 0 14:12:00 14:12:53 14:13:05 14:18:00 14:18:54 14:19:27
delta t 00:00:53 00:00:54
delta t [s] 53 54
Cycle 4 t = 0 14:26:00 14:26:53 14:27:40 14:33:00 14:33:52 14:34:40
delta t 00:00:53 00:00:52
delta t [s] 53 52
80%
0.0 0.9 1.0 1.0 0.1 0.0
0.00 62.07 68.97 68.97 6.90 0.00
Cycle 1 t = 0 09:41:00 09:41:40 09:42:30 09:48:00 09:48:54 09:48:40
delta t 00:00:40 00:00:54
delta t [s] 40 54
Cycle 2 t = 0 09:54:00 09:54:58 09:55:13 10:00:00 10:00:56 10:01:41
delta t 00:00:58 00:00:56
delta t [s] 58 56
Cycle 3 t = 0 14:12:00 14:12:53 14:13:05 14:18:00 14:18:56 14:19:43
delta t 00:00:53 00:00:56
delta t [s] 53 56
Cycle 4 t = 0 14:26:00 14:26:53 14:27:40 14:33:00 14:33:57 14:34:40
delta t 00:00:53 00:00:57
delta t [s] 53 57
Rise
Fall
Concentration
68.97
Device 2
Device 1
Rise
Fall
Concentration
68.97