TÜV Rheinland Energie und Umwelt GmbH
Air Pollution Control
Report on the performance testing of the Serinus 30 ambient air quality moni-
toring system manufactured by Ecotech Pty Ltd measuring CO,
Report no.: 936/21221977/D_EN
Page 13 of 253
EP_Ecotech_Serinus30_en.docx
1.3 Summary of test results
Performance criterion Minimum requirement Test result Com-
pliance
Page
4 Requirements on the instrument design
4.1 General requirements
4.1.1 Measured value display Shall be available.
The measuring system is fitted with a
measured value display.
yes 36
4.1.2 Easy maintenance
Necessary maintenance of the
measuring system should be
possible without larger effort, if
possible from outside.
Maintenan
ce can be carried out with
usual tools in a reasonable time and
from the outside.
yes 37
4.1.3 Functional check If the operation or the function-
al check require particular in-
struments, they shall be con-
sidered as part of the measur-
ing system and be applied in
the corresponding sub-tests
and included in the assess-
ment.
The tested instrument does not have
an internal device for functional
checks.
not ap-
plicable
38
4.1.4 Set-up times and warm-
up times
The set-up times and warm-up
times shall be specified in the
instruction manuals.
Set-up times and warm-
up times were
determined.
yes 39
4.1.5 Instrument design
The instruction manual shall
include specifications of the
manufacturer regarding the de-
sign of the system.
The specifications of the manual with
regard to instrument design are com-
plete and correct.
yes 40
4.1.6 Unintended adjustment
It shall be possible to secure
the adjustment of the measur-
ing system against illicit or un-
intended adjustment during
operation.
The
measuring system itself is not
protected against the unintended or
unauthorised adjustment of instru-
ment parameters. It has to be operat-
ed in a lockable measuring container.
no 41
4.1.7 Data output
The output signals shall be
provided digitally and/or as an-
alogue signals.
Measured signals are provided in
analogue (020 mA, 220 mA, 420
mA or 0
5 V) and digital form (via
TCP/IP, RS 232, USB; Bluetooth).
yes 42