Report on the performance testing of the Serinus 30 ambient air quality
monitoring system manufactured by Ecotech Pty Ltd measuring CO, Report
EP_Ecotech_Serinus30_en.docx
4.3 Field test
The field test was carried out with two complete and identical Serinus 30 measuring systems
and lasted from 04 July 2013 04 October 2013. The measuring systems under test were
identical to those used during laboratory testing. The serial numbers are as follows:
System 1: SN 12-1183
System 2: SN 13-0093
The following performance criteria were tested during the field test:
• Long-term drift
• Maintenance interval
• Availability
• Reproducibility standard deviation under field conditions