TÜV Rheinland Energie und Umwelt GmbH
Luftreinhaltung
Seite 88 von 253
Report on the performance testing of the Serinus 30 ambient air quality
monitoring system manufactured by Ecotech Pty Ltd measuring CO, Report
EP_Ecotech_Serinus30_en.docx
7.4 Evaluation
The following sensitivity coefficients for the influence of sample gas pressure were deter-
mined.
Table 14: Sensitivity coefficient to sample gas pressure
7.5 Assessment
For system 1, the sensitivity coefficient to sample gas pressure is 0.04 µmol/mol/kPa.
For system 2, the sensitivity coefficient to sample gas pressure is 0.03 µmol/mol/kPa.
Does this comply with the performance criterion? yes
7.6 Detailed presentation of test results
Table 15: Individual test results for the influence of changes in sample gas pressure
Sensitivity coeff. sample gas pressure b
gp
[µmol/mol/kPa]
Device 1 Device 2
Time Pressure [kPa] C
oncentration [µmol/mol] [µmol/mol]
10:33:00 80 64.66 64.05 64.89
10:34:00 80 64.66 64.04 64.85
10:35:00 80 64.66 64.03 64.84
10:12:00 110 64.66 65.
28 64.02
10:13:00 110 64.66 65.27 64.02
10:14:00 110 64.66 65.27 64.02