TÜV Rheinland Energie und Umwelt GmbH
Luftreinhaltung
Seite 86 von 253
Report on the performance testing of the Serinus 30 ambient air quality
monitoring system manufactured by Ecotech Pty Ltd measuring CO, Report
no.: 936/21221977/D_EN
EP_Ecotech_Serinus30_en.docx
Table 13: Individual results of the “lack of fit” test
Device 1 [µmol/mol] Device 2 [µmol/mol]
Time Level [%]
Actual value y
i
Set value x
i
Actual value y
i
Set value x
i
10:45:56 80 69.45 68.97 68.55 68.97
10:46:51 80 69.61 68.97 68.82 68.97
10:47:46 80 69.61 68.97 68.86 68.97
10:48:41 80 69.61 68.97 68.86 68.97
10:49:36 80 69.61 68.97 68.86 68.97
11:01:01 40 34.34 34.48 34.88 34.48
11:01:56 40 34.32 34.48 34.81 34.48
11:02:51 40 34.32 34.48 34.81 34.48
11:03:46 40 34.32 34.48 34.81 34.48
11:04:41 40 34.32 34.48 34.81 34.48
11:16:01 0 -0.05 0.00 0.03 0.00
11:16:56 0 -0.10 0.00 -0.05 0.00
11:17:51 0 -0.11 0.00 -0.05 0.00
11:18:46 0 -0.11 0.00 -0.05 0.00
11:19:41 0 -0.11 0.00 -0.05 0.00
11:47:01 60 51.78 51.72 51.73 51.72
11:47:56 60 51.78 51.72 51.73 51.72
11:48:51 60 51.78 51.72 51.73 51.72
11:49:46 60 51.78 51.72 51.76 51.72
11:50:41 60 51.78 51.72 51.78 51.72
12:02:01 20 17.24 17.24 17.40 17.24
12:02:56 20 17.24 17.24 17.40 17.24
12:03:51 20 17.24 17.24 17.40 17.24
12:04:46 20 17.24 17.24 17.40 17.24
12:05:41 20 17.24 17.24 17.40 17.24
12:17:01 95 82.98 81.90 81.41 81.90
12:17:56 95 82.95 81.90 81.41 81.90
12:18:51 95 82.92 81.90 81.41 81.90
12:19:46 95 82.92 81.90 81.41 81.90
12:20:41 95 82.92 81.90 81.41 81.90