Luftreinhaltung
Report on the performance testing of the Serinus 30 ambient air quality moni-
toring system manufactured by Ecotech Pty Ltd measuring CO, Report no.:
936/21221977/D_EN
Page 49 of 253
EP_Ecotech_Serinus30_en.docx
6.1 5.2.6 Switch-over
Switch-over between measurement and functional check and/or calibration shall be
possible telemetrically by computer control or manual intervention.
6.2 Equipment
No additional equipment is required to test this performance criterion.
6.3 Testing
The measuring system may be monitored or controlled via the control panel of the analyser
or telemetrically via remote control.
6.4 Evaluation
All control functions which do not require direct on-site intervention may be performed by op-
erating staff on-site or telemetrically via remote control.
6.5 Assessment
In general, all necessary tasks related to functional checks may be performed directly on-site
or monitored telemetrically using the remote control functions.
Does this comply with the performance criterion? yes
6.6 Detailed presentation of test results
Not applicable in this instance.