TÜV Rheinland Energie und Umwelt GmbH
Luftreinhaltung
Seite 90 von 253
Report on the performance testing of the Serinus 30 ambient air quality
monitoring system manufactured by Ecotech Pty Ltd measuring CO, Report
EP_Ecotech_Serinus30_en.docx
7.4 Evaluation
Table 16: Sensitivity coefficient to sample gas temperature
7.5 Assessment
For system 1, the sensitivity coefficient to sample gas temperature is 0.00 µmol/mol/K).
For system 2, the sensitivity coefficient to sample gas temperature is 0.00 µmol/mol/K).
Does this comply with the performance criterion? yes
7.6 Detailed presentation of test results
Table 17: Individual values obtained from the determination of the influence of sample gas
temperature for NO
Sensitivity coeff. sample gas pressure b
gt
[µmol/mol/K]
Device 1 Device 2
Time Tem
p [ºC] Concentration [µmol/mol] [µmol/mol]
13:37:00 0 64.66 66.11 64.93
13:40:00 0 64.66 66.17 64.93
13:43:00 0 64.66 66.17 64.93
15:10:00 30 64.66 66.01 64.82
15:15:00 30 64.66 66.01 64.82
15:19:00 30 64.66 66.01 64.82