Report on the performance testing of the Serinus 30 ambient air quality
monitoring system manufactured by Ecotech Pty Ltd measuring CO, Report
no.: 936/21221977/D_EN
EP_Ecotech_Serinus30_en.docx
6.1 5.2.9 Instrument software
The version of the instrument software to be tested shall be displayed during switch-
on of the measuring system. The test institute shall be informed on changes in the
instrument software, which have influence on the performance of the measuring sys-
tem.
6.2 Equipment
No additional equipment is required to test this performance criterion.
6.3 Testing
It was verified whether the instrument displays its current software version upon switch-on.
The instrument manufacturer was advised to inform the test institute on any changes to the
instrument software.
6.4 Evaluation
The current software version is displayed upon switch-on of the instrument. It may also be
accessed at any time in the “configuration” menu.
The test was performed while software version 2.09.0005 was in use.
6.5 Assessment
The instrument software version is indicated in the display. Changes to the software will be
communicated to the test institute.
Does this comply with the performance criterion? yes
6.6 Detailed presentation of test results
Figure 8: Display of the software version (2.09.0005) on the start screen