TÜV Rheinland Energie und Umwelt GmbH
Luftreinhaltung
Seite 66 von 253
Report on the performance testing of the Serinus 30 ambient air quality
monitoring system manufactured by Ecotech Pty Ltd measuring CO, Report
no.: 936/21221977/D_EN
EP_Ecotech_Serinus30_en.docx
6.1 5.3.14 Response time
The response time (rise) of the measuring system shall not exceed 180 s.
The response time (fall) of the measuring system shall not exceed 180 s.
The difference between the response time (rise) and response time (fall) of the
measuring system shall not exceed 10 % of response time (rise) or 10 s, whatever
value is larger.
6.2 Equipment
Not applicable here.
6.3 Testing
Performance and evaluation of the steps taken to determine the response time are in line
with the requirements stipulated in DIN EN 14626 (2012). The reader is therefore referred to
section 7.1 8.4.3 Response time.
6.4 Evaluation
Please refer to section 7.1 8.4.3 Response time.
6.5 Assessment
Please refer to section 7.1 8.4.3 Response time.
Does this comply with the performance criterion? yes
6.6 Detailed presentation of test results
Not applicable in this instance.