TÜV Rheinland Energie und Umwelt GmbH
Luftreinhaltung
Seite 38 von 253
Report on the performance testing of the Serinus 30 ambient air quality
monitoring system manufactured by Ecotech Pty Ltd measuring CO, Report
no.: 936/21221977/D_EN
EP_Ecotech_Serinus30_en.docx
6.1 4.1.3 Functional check
If the operation or the functional check of the measuring system require particular
instruments, they shall be considered as part of the measuring system and be ap-
plied in the corresponding sub-tests and included in the assessment.
Test gas units included in the measuring system shall indicate their operational
readiness to the measuring system by a status signal and shall provide direct as
well as remote control via the measuring system.
6.2 Equipment
Manual
6.3 Testing
The tested instrument does not have an internal device for functional checks. The operation-
al status of the AMS is continually monitored and potential problems are displayed via an ar-
ray of different error messages.
The functional check was performed with external test gases.
6.4 Evaluation
The tested instrument does not have an internal device for functional checks. The operation-
al status of the AMS is continually monitored and potential problems are displayed via an ar-
ray of different error messages.
It is possible to perform external zero point and span point checks by means of test gases.
6.5 Assessment
The tested instrument does not have an internal device for functional checks.
Does this comply with the performance criterion? not applicable
6.6 Detailed presentation of test results
Not applicable in this instance.