EasyManuals Logo

Ecotech Serinus 30 User Manual

Ecotech Serinus 30
254 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #77 background imageLoading...
Page #77 background image
TÜV Rheinland Energie und Umwelt GmbH
Luftreinhaltung
Report on the performance testing of the Serinus 30 ambient air quality moni-
toring system manufactured by Ecotech Pty Ltd measuring CO, Report no.:
936/21221977/D_EN
Page 77 of 253
EP_Ecotech_Serinus30_en.docx
Table 8: Individual test results for the short-term drift (initial values)
Device 1 Device 2 Device 1 Device 2
Ti
me [µmol/mol] [µmol/mol] Time [µmol/mol] [µmol/mol]
06:37:00 -0.1 -0.1 07:08:00 65.2 64.5
06:38:00 -0.1 -0.1 07:09:00 65.2 64.5
06:39:00 -0.1 -0.1 07:10:00 65.2 64.5
06:40:00 -0.1 -0.1 07:11:00 65.2 64.5
06:41:00 -0.1 -0.1 07:12:00 65.2 64.5
06:42:00 -0.1 -0.1 07:13:00 65.2 64.5
06:43:00 -0.1 -0.1 07:14:00 65.2 64.5
06:44:00 -0.1 -0.1 07:15:00 65.2 64.5
06:45:00 -0.1 0.0 07:16:00 65.2 64.5
06:46:00 -0.1 0.0 07:17:00 65.2 64.5
06:47:00 -0.1 0.0 07:18:00 65.2 64.6
06:48:00 -0.1 0.0 07:19:00 65.2 64.6
06:49:00 -0.1 0.0 07:20:00 65.2 64.6
06:50:00 -0.1 0.0 07:21:00 65.2 64.6
06:51:00 -0.1 0.0 07:22:00 65.2 64.6
06:52:00 -0.1 0.0 07:23:00 65.3 64.6
06:53:00 -0.1 0.0 07:24:00 65.3 64.6
06:54:00 -0.1 0.0 07:25:00 65.3 64.7
06:55:00 -0.1 0.0 07:26:00 65.3 64.7
06:56:00 -0.1 0.0 07:27:00 65.3 64.7
Average -0.1 0.0 Average 65.2 64.6
Start values
Start values
Zero level
Span level

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Ecotech Serinus 30 and is the answer not in the manual?

Ecotech Serinus 30 Specifications

General IconGeneral
BrandEcotech
ModelSerinus 30
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals