TÜV Rheinland Energie und Umwelt GmbH
Luftreinhaltung
Report on the performance testing of the Serinus 30 ambient air quality moni-
toring system manufactured by Ecotech Pty Ltd measuring CO, Report no.:
936/21221977/D_EN
Page 45 of 253
EP_Ecotech_Serinus30_en.docx
6.1 5.2.2 Measuring range
The upper limit of measurement of the measuring systems shall be greater or equal
to the upper limit of the certification range.
6.2 Equipment
No additional equipment is required to test this criterion.
6.3 Testing
It was determined whether the upper limit of the measuring range was greater or equal to the
upper limit of the certification range.
6.4 Evaluation
In principle, the measuring system allows for measuring ranges from max. 0 – 200 ppm.
Possible measuring range: 200 ppm
Upper limit of the certification range for CO: 100 mg/m³
6.5 Assessment
By default the measuring range is set to 0 100 mg/m³ for CO. Other measuring ranges of
max. 0 200 ppm are possible.
The upper limit of the measuring range is larger than the respective upper limit of the certifi-
cation range.
Does this comply with the performance criterion? yes
6.6 Detailed presentation of test results
Not applicable in this instance.