TÜV Rheinland Energie und Umwelt GmbH
Luftreinhaltung
Seite 84 von 253
Report on the performance testing of the Serinus 30 ambient air quality
monitoring system manufactured by Ecotech Pty Ltd measuring CO, Report
EP_Ecotech_Serinus30_en.docx
Table 12: Residues of the analytical function for CO
Figure 10: Function established from group averages for system 1, component CO
Residual at zero r
z
[µmol/mol]
Largest value of the relative residuals r
m a x
[%]
0.0
-0.9
-0.6
0.5
0.3
y = 1.0131x - 0.3105
-4.00
-3.00
-2.00
-1.00
0.00
1.00
2.00
3.00
4.00
-10
0
10
20
30
40
50
60
70
80
90
0 10 20 30 40 50 60 70 80 90
Actual values [µmol/mol]
Set values [µmol/mol]
Lack of fit in [µmol/mol]
Test points device 1
Residues Device 1
Linear (Test points device 1)