TÜV Rheinland Energie und Umwelt GmbH
Luftreinhaltung
Report on the performance testing of the Serinus 30 ambient air quality moni-
toring system manufactured by Ecotech Pty Ltd measuring CO, Report no.:
936/21221977/D_EN
EP_Ecotech_Serinus30_en.docx
7.5 Assessment
The sensitivity coefficient of electrical voltage b
v
does not exceed the performance criteria of
max. 0.30 µmol/mol/V stipulated in Standard DIN EN 14626 at any point. For both systems,
the highest value b
v
is used for the purpose of evaluating uncertainty. For system 1 it is
0.001 µmol/mol/V and for system 2 it is 0.001 µmol/mol/V.
Does this comply with the performance criterion? yes
7.6 Detailed presentation of test results
Table 21: Individual results for the tests of the sensitivity coefficient to electrical voltage
Device 1 Device 2
Time Vo
ltage [V] Concentration [µmol/mol] [µmol/mol]
09:55:00
198
0 -0.16 0.00
09:56:00
198
0 -0.16 0.00
09:57:00
198
0 -0.16 -0.05
-0.16 -0.02
10:03:00
264
0 -0.22 -0.05
10:04:00
264
0 -0.22 0.00
-0.22 -0.04
10:15:00
198
64.50 64.88 64.18
10:16:00
198
64.50 64.88 64.18
10:17:00
198
64.50 64.88 64.18
64.88 64.18
10:22:00
264
64.50 64.93 64.23
10:23:00
264
64.50 64.93 64.23
10:24:00
264
64.50 64.93 64.23
64.93 64.23