TÜV Rheinland Energie und Umwelt GmbH
Luftreinhaltung
Report on the performance testing of the Serinus 30 ambient air quality moni-
toring system manufactured by Ecotech Pty Ltd measuring CO, Report no.:
936/21221977/D_EN
EP_Ecotech_Serinus30_en.docx
7.4 Evaluation
Table 5: Response times of the two Serinus 30 measuring systems for CO
For system 1 this results in a maximum t
r
of 50 s, a maximum t
f
of 52 s and a t
d
of -2 s for
CO.
For system 2 this results in a maximum t
r
of 51 s, a maximum t
f
of 56 s and a t
d
of -5 s for
CO.
7.5 Assessment
The maximum permissible response time of 180 s is exceeded at no time. The maximum re-
sponse time determined is 52 s for system 1 and 56 s for system 2.
Does this comply with the performance criterion? yes