TÜV Rheinland Energie und Umwelt GmbH
Luftreinhaltung
Report on the performance testing of the Serinus 30 ambient air quality moni-
toring system manufactured by Ecotech Pty Ltd measuring CO, Report no.:
936/21221977/D_EN
Page 117 of 253
EP_Ecotech_Serinus30_en.docx
8. Recommendations for use
Tasks during the period of unattended operation (4 weeks)
The following checks are required on a regular basis:
Regular visual inspections / telemetric monitoring
Analyser status
Error warnings
Replacement of the Teflon filter in the sample gas inlet
Zero and span point checks with appropriate test gases
For all intents and purposes the manufacturer’s instructions shall be considered.
Further details are provided in the manual.
Immissionsschutz/Luftreinhaltung
______________________________
______________________________
Dipl.-Ing. Guido Baum
Dipl.-Ing. Karsten Pletscher
Cologne, 08 October 2013