Report on the performance testing of the Serinus 30 ambient air quality
monitoring system manufactured by Ecotech Pty Ltd measuring CO, Report
EP_Ecotech_Serinus30_en.docx
6.1 5.3.10 Averaging effect
For gaseous components the measuring system shall allow the formation of hourly
averages.
The averaging effect shall not exceed the requirements of Table 2 of VDI 4202,
Sheet 1 (September 2010).
The averaging effect shall not exceed 7 % of the measured value.
6.2 Equipment
Not applicable here.
6.3 Testing
Performance and evaluation of the steps taken to determine the averaging effect are in line
with the requirements stipulated in DIN EN 14626 (2012). The reader is therefore referred to
section 7.1 8.4.12 Averaging test.
6.4 Evaluation
Please see section 7.1 8.4.12 Averaging test.
6.5 Assessment
Please see section 7.1 8.4.12 Averaging test
Does this comply with the performance criterion? yes
6.6 Detailed presentation of test results