TÜV Rheinland Energie und Umwelt GmbH  
Luftreinhaltung  
Report on the performance testing of the Serinus 30 ambient air quality moni-
toring system manufactured by Ecotech Pty Ltd measuring CO, Report no.: 
936/21221977/D_EN 
 
 
 
EP_Ecotech_Serinus30_en.docx 
 
7.4    Evaluation 
Table 27:  Results for the long term drift at zero for the component CO 
 
Table 28:  Results for the long term drift at span point for the component CO 
 
Device 1 Device 2
[µmol/mol] [µmol/mol]
C
Z,0
    04.07.2013 0.00 0.00
C
Z,1
    22.07.2013 -0.04 -0.17
D
L,Z
    22.07.2013
-0.04 -0.17
C
Z,1
    02.08.2013 -0.14 0.09
D
L,Z
    02.08.2013
-0.14 0.09
C
Z,1
    16.08.2013 -0.23 0.10
D
L,Z
    16.08.2013
-0.23 0.10
C
Z,1
    02.09.2013 -0.21 -0.47
D
L,Z
    02.09.2013
-0.21 -0.47
C
Z,1
    16.09.2013 0.03 0.36
C
Z,1
    30.09.2013 0.06 -0.27
D
L,Z
    30.09.2013
0.06 -0.27
C
Z,1
    04.10.2013 0.15 0.03
Device 1 Device 2
[µmol/mol] [µmol/mol]
C
S,0
    04.07.2013 64.6 64.6
C
S,1
    22.07.2013 64.8 64.5
D
L,S
    22.07.2013
0.37% 0.13%
C
S,1
    02.08.2013 64.7 64.6
D
L,S 
  02.08.2013
0.47% -0.13%
C
S,1
    16.08.2013 64.8 64.3
D
L,S
    16.08.2013
0.64% -0.59%
C
S,1
    02.09.2013 64.6 64.6
D
L,S
    02.09.2013
0.44% 0.66%
C
S,1
    16.09.2013 64.7 64.4
D
L,S
    16.09.2013
0.13% -0.90%
C
S,1
    30.09.2013 64.8 64.3
D
L,S
    30.09.2013
0.26% -0.05%
C
S,1
    04.10.2013 64.7 64.4
D
L,S
    04.10.2013
0.01% -0.37%