TÜV Rheinland Energie und Umwelt GmbH
Luftreinhaltung
Report on the performance testing of the Serinus 30 ambient air quality moni-
toring system manufactured by Ecotech Pty Ltd measuring CO, Report no.:
936/21221977/D_EN
Page 105 of 253
EP_Ecotech_Serinus30_en.docx
7.4 Evaluation
Table 27: Results for the long term drift at zero for the component CO
Table 28: Results for the long term drift at span point for the component CO
Device 1 Device 2
[µmol/mol] [µmol/mol]
C
Z,0
04.07.2013 0.00 0.00
C
Z,1
22.07.2013 -0.04 -0.17
D
L,Z
22.07.2013
-0.04 -0.17
C
Z,1
02.08.2013 -0.14 0.09
D
L,Z
02.08.2013
-0.14 0.09
C
Z,1
16.08.2013 -0.23 0.10
D
L,Z
16.08.2013
-0.23 0.10
C
Z,1
02.09.2013 -0.21 -0.47
D
L,Z
02.09.2013
-0.21 -0.47
C
Z,1
16.09.2013 0.03 0.36
D
L,Z
16.09.2013
0.03 0.36
C
Z,1
30.09.2013 0.06 -0.27
D
L,Z
30.09.2013
0.06 -0.27
C
Z,1
04.10.2013 0.15 0.03
D
L,Z
04.10.2013
0.15 0.03
Device 1 Device 2
[µmol/mol] [µmol/mol]
C
S,0
04.07.2013 64.6 64.6
C
S,1
22.07.2013 64.8 64.5
D
L,S
22.07.2013
0.37% 0.13%
C
S,1
02.08.2013 64.7 64.6
D
L,S
02.08.2013
0.47% -0.13%
C
S,1
16.08.2013 64.8 64.3
D
L,S
16.08.2013
0.64% -0.59%
C
S,1
02.09.2013 64.6 64.6
D
L,S
02.09.2013
0.44% 0.66%
C
S,1
16.09.2013 64.7 64.4
D
L,S
16.09.2013
0.13% -0.90%
C
S,1
30.09.2013 64.8 64.3
D
L,S
30.09.2013
0.26% -0.05%
C
S,1
04.10.2013 64.7 64.4
D
L,S
04.10.2013
0.01% -0.37%

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Ecotech Serinus 30 and is the answer not in the manual?

Ecotech Serinus 30 Specifications

General IconGeneral
BrandEcotech
ModelSerinus 30
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals