TÜV Rheinland Energie und Umwelt GmbH
Luftreinhaltung
Report on the performance testing of the Serinus 30 ambient air quality moni-
toring system manufactured by Ecotech Pty Ltd measuring CO, Report no.:
936/21221977/D_EN
Page 111 of 253
EP_Ecotech_Serinus30_en.docx
Evaluation
The down times during the field test are listed in Table 31.
Table 31: Availability of the Serinus 30 analyser
System 1 System 2
Time in the field h 2244 2244
Down time h 0 0
Maintenance h 12 12
Total operating time h 2232 2232
Total operating time incl.
maintenance
2244 2244
Availability % 100 100
Maintenance times result from the daily test gas feeding for the purpose of determining the
drift behaviour and the maintenance interval as well as times required to change the internal
Teflon filters in the sample gas line.
7.5 Assessment
The availability is 100 %. This complies with the requirements of Standard DIN EN 14626.
Does this comply with the performance criterion? yes
7.6 Detailed presentation of test results
Not applicable in this instance.