TÜV Rheinland Energie und Umwelt GmbH
Luftreinhaltung
Seite 72 von 253
Report on the performance testing of the Serinus 30 ambient air quality
monitoring system manufactured by Ecotech Pty Ltd measuring CO, Report
no.: 936/21221977/D_EN
EP_Ecotech_Serinus30_en.docx
Figure 9: Diagram illustrating the response time
7.3 Testing
The test was performed in accordance with the requirements on testing as stipulated in DIN
EN 14626. Data were recorded using a Yokogawa DX2000 data logger with its averaging
time set to 1 s.
Key
A analyser response
1 lag time
2 rise time
3 response time (rise)
4 fall time
5 response time (fall)