Bit num-
ber
Status of bit (in example)
Indicaon
7 0 Speed is below 50% of full rot
aonal speed
8 0 The pump is not in parallel control mode
9 0 The pump is not in serial control mode
10 0
There is no controller internal soware mis-
match
11 0
Controller passed internal conguraon and cal-
ibraon operaon
12 0 The mer has not med out
13 0 Overspeed and overcurrent trip not acvated
14 0
Pump internal temperature measurement sys-
tem is ne
15 0
Serial enable has not become inacve during se-
rial control
11.6 Service informaon
If using the serial communicaons link, addional informaon about the pump, such as
pump type and internal controller soware versions, can be accessed. This informaon
is parcularly useful for service personnel to determine the model of the pump.
Send the following query to nd out pump type:
Command
? s 8 5 1 cr
The reply is as follows, where String 1 is the pump type, String 2 is the DSP sowar
e
version number and String 3 is the designated full speed of the pump (in
revoluons per second):
Reply
= s 8 5 1 sp String 1 ; String 2 ; String 3 cr
Send the following query to nd out the PIC soware version:
Command
? s 8 6 8 cr
The reply is as follows, where String 1 is the PIC sowar
e version number:
Reply
= s 8 6 8 sp String 1 cr
08/2021 - ©Edwards Limited
Page 104B8J200880_D
300812671_002_C3
B8J200880_D - Fault nding